Appendix3 List of default settings
A25
Appendix
Touch current (Enclosure - Line)
measurement manual manual
–
condition L side L side
Selected line internal internal
filter ON1_U2 ON1_U2
target current ACDC ACDC
range AUTO AUTO
allowable value : upper limit : normal – –
allowable value : upper limit : fault 3.500 mA 250.0
A
allowable value : lower limit :
ON/OFF : normal
––
allowable value : lower limit :
ON/OFF : fault
OFF OFF
Free current
measurement manual manual –
polarity
positive
phase
positive
phase
–
equipment status normal normal –
filter ON1_U2 ON1_U2 ON1_U2
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value : upper limit : normal 3.500 mA 250.0
A250.0 A
allowable value : upper limit : fault 3.500 mA 250.0
A–
allowable value : lower limit :
ON/OFF : normal
OFF OFF OFF
allowable value : lower limit :
ON/OFF : fault
OFF OFF –
default setting
Leakage current mode item Class I Class II Internally
powered
*
( ) is the value after DC switching