Appendix3 List of default settings
A26
Network D/E/F
After selecting network, class, applied part and leakage current mode
default setting
Leakage current mode item Class I Class II Internally
powered
Earth leakage current
measurement manual
––
polarity positive phase
equipment status normal
filter –
target current ACDC
range AUTO
allowable value : upper limit : normal 3.500 mA
allowable value : upper limit : fault 3.500 mA
allowable value : lower limit :
ON/OFF : normal
OFF
allowable value : lower limit :
ON/OFF : fault
OFF
Enclosure - Earth
leakage current
measurement manual manual –
polarity positive phase positive phase –
equipment status normal normal –
filter – – –
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value : upper limit : normal 3.500 mA 3.500 mA 3.500 mA
allowable value : upper limit : fault 3.500 mA 3.500 mA –
allowable value : lower limit :
ON/OFF : normal
OFF OFF OFF
allowable value : Flower limit :
ON/OFF : fault
OFF OFF –
Enclosure - Enclosure
leakage current
measurement manual manual –
polarity positive phase positive phase –
equipment status normal normal –
filter – – –
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value : upper limit : normal 3.500 mA 3.500 mA 3.500 mA
allowable value : upper limit : fault 3.500 mA 3.500 mA –
allowable value : lower limit :
ON/OFF : normal
OFF OFF OFF
allowable value : lower limit :
ON/OFF : fault
OFF OFF –
Enclosure - Line
leakage current
measurement manual manual
–
condition L side L side
Selected line internal internal
filter OFF OFF
target current ACDC ACDC
range AUTO AUTO
allowable value : upper limit : normal – –
allowable value : upper limit : fault 3.500 mA 3.500 mA
allowable value : lower limit :
ON/OFF : normal
––
allowable value : lower limit :
ON/OFF : fault
OFF OFF
*
( ) is the value after DC switching