Appendix3 List of default settings
A27
Appendix
Network G
After selecting network, class, applied part and leakage current mode
default setting
Leakage current mode item Class I Class II Internally
powered
Earth leakage current
measurement manual
––
polarity
positive
phase
equipment status normal
filter OFF
target current ACDC
range AUTO
allowable value : upper limit : normal 500.0
A
allowable value : upper limit : fault 3.500 mA
allowable value : lower limit :
ON/OFF : normal
OFF
allowable value : lower limit :
ON/OFF : fault
OFF
Touch current (Enclosure - Earth)
measurement manual manual
polarity polarity
positive
phase
positive
phase
equipment status normal normal
filter OFF OFF OFF
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value : upper limit : normal 500.0
A500.0 A500.0 A
allowable value : upper limit : fault 3.500 mA 3.500 mA
allowable value : lower limit :
ON/OFF : normal
OFF OFF OFF
allowable value : lower limit :
ON/OFF : fault
OFF OFF
Touch current
(Enclosure - Enclosure)
measurement manual manual
polarity
positive
phase
positive
phase
equipment status normal normal
filter OFF OFF OFF
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value : upper limit : normal 500.0
A500.0 A500.0 A
allowable value : upper limit : fault 3.500 mA 3.500 mA
allowable value : lower limit :
ON/OFF : normal
OFF OFF OFF
allowable value : lower limit :
ON/OFF : fault
OFF OFF
*
( ) is the value after DC switching