4-47 
4.10.2    Makeup Reservoir replacement   
1 Drain makeup from the makeup Reservoir. Execute “Reservoir Makeup Drainage” and 
  drain the makeup from the makeup reservoir. 
[Caution] In case that the makeup doesn't come out from the makeup reservoir even 
when  "Reservoir Makeup Drainage" is executed, which may be caused by malfunction 
of the part such as the solenoid valve or the pump, remove the makeup reservoir in the 
step [5] below and drain the makeup into a beaker out of the makeup reservoir's opening 
for attaching the sensor block. At the makeup drainage, put wiping paper under the 
beaker for possible makeup dripping. If there is still makeup remaining in the makeup 
cartridge bottle, do NOT lift the lever handle. 
 
2 Turn off the power supply. 
 
3 Perform the procedures described in “4.1.2 Access to the parts on the front side of the 
  equipment” and pull the reservoir unit forward. 
[Caution] Be careful not to catch the tubes. 
 
4 Remove the Sensor block from the Makeup reservoir. 
[Caution] When removing the Sensor block, be careful not to have an impact on the 
sensor nor vibration 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
5 Remove the Fixing screws (2 positions) and remove the Makeup Reservoir. 
 
6 Lift the lever handle and remove the Makeup Cartridge.