2-18
2.3.6 Heating unit confirmation
The confirmation of heating unit operation is performed.
Activate/Deactivate the heating unit.
The ambient temperature and heating unit temperature are displayed.
Perform this confirmation in a state other than "Ready" state.
1 Press Adjustment/operational checkout on the maintenance menu.
If Adjustment/operational checkout is not displayed on the maintenance menu, set the
service personnel maintenance mode on the login user selection screen.
The Adjustment/operational checkout screen is displayed.
For information on changing to the service personnel maintenance mode, refer
to “2.1 Setting of service personnel maintenance mode.”
Manual
Start
up
HOME
Back
Adjust./oper. checkout
Excitation
adjustment
Charge voltage
Confirmation N1
Phase margin test
Heating unit
confirmation
Charge voltage
Confirmation N2
Auto-phase gain
adjustment
Level sensor
confirmation
[Stop ]
2018.07.07 12:45
Com=0