Section VII 
A3TP6:  > 
+6 
V 
A3TP5:  0 V ± 
.1 
V 
A3TP8:  + 9.5 V ±  .5 V 
Collector 
A3Ql6: 
- 9.9 V ± 
.1 
V 
3.  Reposition: 
RESOLUTION BANDWIDTII  . . . . .  1 Hz 
4.  Connect  an  oscilloscope  to  the  collector 
of 
A3Ql 
l. 
Wait  5 seconds. The voltage should be 
a 
TTL 
HIGH 
(;;;;.. 
2 V de). 
5.  Adjust 
A3Rl4 
fully 
CCW*. 
6.  While  watching  the  oscilloscope  clock 
Sl. 
The 
oscilloscope  should  indicate  a  TTL 
LOW 
( < 
.8 
V)  for  a  few  seconds  and then return 
HIGH. The logic state should be 011. 
v. 
State 011 
1. 
Check the following levels: 
A3TP5:  OV 
de± 
.1 
V 
A3TP6:  0 V de ± 
.1 
V 
Collector 
of 
A3Ql6: 
- 9.9 V ± 
.1 
V 
A3TP8:  TTL 
LOW 
( <  .8 V) 
2.  Reposition: 
RESOLUTION BANDWIDTII  . . .  100 Hz 
3.  Clock 
Sl 
and the control logic should 
go 
to 
state 001. 
w.  StateOOl 
1.  Clock 
Sl 
and observe no change 
of 
state. 
2.  Check the following levels: 
Collector A3Q16:  - 9.9 V ± 
.1 
V 
A3TP5: 
<-7 
V 
A3TP8:  TTL 
LOW 
( < 
.8 
V) 
3.  Adjust 
A3Rl4 
fully 
CW*. 
Clock 
SI, 
and  the 
control logic should 
go 
to 
State 101. 
x.  Adjust  R14  so  that the  voltage  at A3TP1 l 
is 
at the 
transition between a plus and minus voltage. 
y.  Remove  all  test  leads  and replace the clock jumper. 
The  3580A  should  sweep  normally.  The  penlift  relay 
should "click" in single  sweep mode  and the 
output 
of 
the 
A3 
RAMP  GENERATOR  (A3TP1)  should  be  + 5  volts 
nominal  for  a  front  panel  display  indication  at the  right 
graticule. 
If 
the LOG  SWEEP mode will 
not 
work, see  the 
A3 
schematic notes. 
Model 3580A 
7-12. 
A7 
Board. 
7-13.  The A 7  Board (03580-66507) is available 
as 
a rebuilt 
exchange  board  (03580-69507)  through  your  local  -hp-
Sales  and  Service  Office.  Many times, however,  the  board 
can be repaired without purchasing an exchange board. The 
following procedure will aid in determining whether the A 7 
board or the analog circuits preceeding 
the~ 
A 7 board are 
at 
fault. 
a.  Connect the 3580A X-AXIS 
output 
on the rear panel 
to 
the X deflection EXT INPUT 
of 
an oscilloscope. A scope 
with  variable  persistance  works  best 
but 
is 
not 
absolutely 
necessary.  Connect  the  3580A  Y-AXIS 
output 
to 
the 
vertical  input 
of 
the scope. This procedure effectively half 
splits the 3580A for troubleshooting purposes. 
1. 
If 
the  signal  seen 
of 
the  scope 
is 
correct and the 
signal  seen  on the 3580A display 
is 
incorrect then 
the problem 
is 
in  the  A7,  A8, 
or 
A13 boards. 
If 
the signal seen  on the scope 
is 
incorrect 
DO 
NOT 
troubleshoot the  A 7 board until 
repiii.rs 
are  made 
to 
preceeding  circuitry.  (See  Functional  Block 
Diagram  in  the  Operating  and  Service  Manual.) 
2. 
If 
the  scope  presentation 
is 
good 
but 
the 3580A 
display 
is 
incorrect, check A 7TP1. 
If 
the presenta-
tion is bad there then troubleshoot the A 7 board, 
otherwise troubleshoot 
A8 
or 
Al3. 
7-14.  Troubleshooting the A 7 board. 
a.  Check A7Q2, Q4, Q6, Q8, and  Q9. 
If 
these parts are 
P/N  1853-0098 replace  all  5 
of 
them with P/N 1853-0010. 
The new type 
is 
much more reliable and 
is 
being used in all 
instruments  with  serial  numbers  above  1415A01276. 
b.  Check A 7TP4 and A 7TP5. They should look similar 
to 
the figure shown below. 
Horiz 2 µ  sec/div 
Vert 
.2 
V/div (with 10:1 probe) 
TP4 
TP5 
The frequency must be  55  K  - 70 kHz! 
If 
the frequency 
is 
off 
check  A8TP9.  The  clock  frequency 
is 
determined 
by 
the A8 board. 
*If 
A3R14 
has a 
black 
casing, 
set 
it 
opposite 
to 
the 
setting given. 
7-4