Model 3580A 
GENERAL 
OPERATING 
INFORMATION 
Section III 
3-23.  For battery  powered instruments (Option 001), the 
common  binding  post 
of 
the  INPUT  connector  (Case 
Ground 
¢" 
) should be connected 
to 
earth ground or 
to 
an  appropriate  system ground. 
If 
a system ground 
is 
used, 
extra 
care 
should 
be 
taken  to  ensure  that 
it 
is 
actually at 
ground potential and 
is 
not 
a voltage source. 
3-24. 
Ground 
Loops. 
3-25.  In the design 
of 
the 3580A, extra care has been taken 
to 
control  internal  ground  currents 
that 
could  produce 
undesirable  responses or degrade  the  accuracy 
of 
low level 
measurements.  Due 
to 
its  wide  dynamic  range  and  high 
sensitivity, however, the 3580A can be  affected by external 
ground  currents  or  "ground  loops"  which  are  normally 
caused by poor grounding. The following paragraphs briefly 
describe  the  common  power-line  ground  loop and outline 
the  steps 
that 
can  be  taken 
to 
minimize  ground  loop 
problems. 
3-26.  Figure  3-5A  shows  the  input  arrangement  for  a 
simple  grounded  measurement. 
Ein 
represents  the  source 
being measured along with any noise associated with 
it 
and 
is 
generally  called the "normal-mode source". 
Rs 
represents 
the source resistance  and the resistance 
of 
the high lead; 
Rg 
represents  the  resistance 
of 
the ground lead. Current from 
Ein  (normal-mode  current)  flows  through 
Rs, 
Z
1 
and 
Rg 
and the instrument responds 
to 
the drop across Z
1
• 
As 
long 
as 
the grounds on 
both 
sides 
of 
Rg 
are identical, extraneous 
currents  cannot  circulate  between the  source  ground and 
the  instrument  ground. 
If, 
however,  the  grounds  are 
different  due 
to 
voltage  drops  in  the  ground  lead  or 
currents induced 
into 
it, 
a new source is developed and the 
measurement  appears 
as 
shown  in  Figure  3-5B.  The  new 
source, 
Ecm 
(the difference  between grounds), 
is 
called the 
"common-mode source" because it 
is 
common 
to 
both 
the 
high  and  ground  lines.  Common-mode  current  can  flow 
Ein 
NORMAL 
MODE 
SOURCE 
Rs 
"HIGH" 
SOURCE 
Rg 
GROUND  LEAD 
RESISTANCE 
SOURCE 
GROUND 
A 
INPUT 
z, 
IMPEDANCE 
INSTRUMENT 
GROUND 
through 
Rg 
or through 
Rs 
and Z
1
•  Since Z
1 
is usually much 
larger than 
Rs 
and since  they are  both in parallel with 
Rg, 
most 
of 
the voltage  across 
Rg 
will  appear across Z
1 
causing 
an error in the amplitude reading. 
3-27.  To  minimize power-line  ground loops, the following 
guidelines should be  observed: 
a. 
Keep input leads as short 
as 
possible. 
b.  Provide good ground connections 
to 
minimize 
Rg. 
c.  Connect the signal source and the 3580A 
to 
the same 
power bus. 
d. 
If 
a removable  ground strap 
is 
provided 
on 
the  signal 
source, float the source 
to 
break the common-mode current 
path. 
e.  Option  001:  Battery  operate  the 3580A; connect a 
separate  ground lead between the common terminal 
of 
the 
3580A INPUT  connector  and  the  ground  terminal 
of 
the 
signal source. 
3-28. 
Measurement 
Configurations. 
3-29.  The  3580A  can 
be 
used in either 
of 
two  measure-
ment  configurations:  open  loop  or  closed  loop.  These 
configurations are illustrated in Figure 3-6. 
3-30. 
Open 
Loop.  In  the  open-loop  configuration,  the 
3580A  functions 
as 
a 
signal 
analyzer  which  divides  the 
input  signal  into  its  various  frequency  components.  The 
amplitudes 
of 
these components are displayed 
as 
a function 
of 
frequency  on  the  CRT.  The  amplitupe 
vs. 
frequency 
display  shows  how energy 
is 
distributed 
as 
a function 
of 
SOURCE 
-:;:-
GROUND 
GROUNDED 
MEASURING  INSTRUMENT 
HIGHr--
-
I  • 
I 
Rg 
j  ' 
• 
I 
j---
----6)---
Ecm 
COMMON 
MODE 
SOURCE 
B 
-:;:-
INSTRUMENT 
GROUND 
3!575-B-2738 
Figure 3-5.  Power  Line Ground  Loop. 
3-5