PERFORMANCE 
TEST 
CARD 
(cont'd) 
BANDWIDTH 
TESTS 
Resolution Bandwidth 
300Hz 
lOOHz 
30Hz 
t()Hz 
Resolution Bandwidth 
3 Hz 
1 
Hz 
NOISE 
LEVEL 
TESTS 
Bandwidth 
300Hz 
30Hz 
30Hz 
30Hz 
1 Hz 
1 Hz 
1 Hz 
NOISE 
SIDEBAND 
TEST 
Lower 3 
dB 
Frequency 
lOkHz 
1 kHz 
1 kHz 
1 kHz 
Lower 60 
dB 
Frequency 
1 kHz 
1 kHz 
Frequency 
lOkHz 
lOkHz 
lOOHz 
1 kHz 
1 kHz 
lOOHz 
lOHz 
Noise  Sidebands  must  be  at least  70 
dB 
below continuous 
wave 
signal,± 10 Hz away. 
SPURIOUS 
RESPONSE 
TEST 
All  non-line-related  spurious  responses  must  be  at  least 
80 
dB 
below a full scale reference. 
LINE-RELATED 
SPURIOUS 
RESPONSE 
TEST 
All 
line-related  spurious  responses  must 
be 
less  than 
- 140 dBV (.1 µV). 
IF FEEDTHRU 
TEST 
IF  Feedthru must  be ·at least  - 70 
dB 
below the full  scale 
reference. 
ZERO 
BEAT 
RESPONSE 
TEST 
The  zero  beat  response  must  be  at  least 
30 
dB 
below the 
full scale reference. 
INPUT 
IMPEDANCE 
TESTS 
Frequency 
Display Indication 
(O 
dB 
= full scale) 
Without 1 Mn  With 1 Mn 
1 kHz 
OdB 
___ 
dB 
Upper 3 
dB 
Frequency 
kHz 
kHz 
kHz 
kHz 
Upper 
60 
dB 
Frequency 
kHz 
kHz 
Noise 
(- 70 
dB= 
full scale) 
dB 
dB 
dB 
dB 
dB 
dB 
dB 
Pass 
Pass 
Pass 
Pass 
Pass 
Test Limits 
10.3 kHz  ± 45 Hz 
1.1 
kHz± 
15 
Hz 
1.030 
kHz± 
4.5 Hz 
1.010 kHz ±  1.5 Hz 
Test Limits 
1.030 
kHz± 
4.5 Hz 
1.010 
kHz± 
1.5 Hz 
Test Limits 
<-130 
dB 
<-
140 
dB 
<-132 
dB 
<-140 
dB 
<-150 
dB 
<-143 
dB 
<-135 
dB 
Fail 
Fail 
Fail 
Fail 
Fail 
Test Limit 
-3dB±ldB