Model 3580A 
GENERAL 
OPERATING 
INFORMATION 
Section III 
Increasing the frequency  span  or decreasing the sweep time 
increases the sweep rate. 
Figure  3-26. 
Log 
Amplitude Vs. 
Log 
Freq. Plot of 
20 
kHz 
Notch  Filter (30 Hz BW). 
3-135.  Optimum 
Sweep 
Rate.  The  optimum sweep rate 
is 
the  maximum  rate  at  which  the  frequency  can 
be 
swept 
without excessively  compressing or skewing the amplitude 
response.  When  the  3580A 
is 
sweeping 
at 
what 
is 
consid-
ered 
to 
be  the  optimum rate, the amplitude compression 
is 
about 
2%. 
3-136.  The  optimum  sweep  rate 
is 
determined  by  the 
response  time 
of 
the  instrument. 
If 
the  response  time 
is 
long,  the  sweep  rate  must 
be 
slow 
so 
that 
the instrument 
can  respond  properly.  The  response  time 
of 
the 3580A 
is 
determined by  the BANDWIDTH and DISPLAY SMOOTH-
ING  settings.  Narrowing  the  bandwidth  or  increasing  the 
display  smoothing  increases  the  response  time  and, there-
fore, decreases the optimum sweep rate. 
3-137. 
Optimum 
Sweep 
Indicator.  The  3580A 
is 
equipped  with 
an 
internal  detector  that  monitors  the 
BANDWIDTH,  DISPLAY  SMOOTHING,  FREQUENCY 
SPAN  and  SWEEP  TIME  control  settings.  When  these 
control  settings  are  such  that  the  sweep  rate  exceeds  the 
optimum  sweep  rate,  the  front  panel  ADJUST  indicator 
illuminates. 
3-138.  To  sweep  at  the  optimum  rate,  first  set  the 
FREQUENCY,  FREQUENCY  SPAN,  BANDWIDTH  and 
DISPLAY  SMOOTHING  controls 
to 
obtain  the  desired 
measurement  parameters.  Then,  starting  with  a  slow 
SWEEP 
TIME  setting,  increase  the  sweep  rate  until  the 
ADJUST  light  first  comes  on.  When  the  ADJUST  light 
comes  on,  rotate  the 
SWEEP 
TIME  control  one  position 
counterclockwise.  The  ADJUST  light  will 
go 
out 
and the 
instrument will sweep at the optimum rate. 
3-139.  Table  3-5  lists  the  optimum SWEEP  TIME  settings 
for  various  FREQ  SPAN,  BANDWIDTH  and  DISPLAY 
SMOOTHING settings. 
3-140. 
For 
closed-loop measurements where  the 
3580A 
is 
used 
as 
a  network  analyzer,  the  optimum  sweep  rate 
is 
determined  by  the  3580A  BANDWIDTH  and  DISPLAY 
SMOOTHING control settings and by the bandwidth 
of 
the 
network under test. During closed-loop measurements, the 
input frequency 
is 
always near the center 
of 
the passband 
and the 
IF 
Filter 
is 
required to respond only 
to 
amplitude 
variations  introduced by the network. For this reason, the 
optimum  sweep  rate  for  closed -loop  measurements 
is 
generally  much  faster  than  it 
is 
for  open-loop measure-
ments.  In  many  closed-loop measurement applications  the 
sweep  rate  can  be  set  20 
to 
25  times  faster  than  the 
optimum rate indicated by the ADJUST light. 
3-141. 
If 
the  optimum  sweep  rate 
is 
not 
limited  by  the 
bandwidth 
of 
the  3580A,  it  may 
be 
limited  by  the 
bandwidth 
of 
the  network  under  test. 
For 
bandpass  and 
low  pass  filters,  a rough  approximation 
of 
optimum sweep 
rate can be made using the following formula: 
BW
2 
R=--
2 
Where: 
R  = optimum sweep rate in Hz/sec 
BW 
=  bandwidth 
of 
network under test 
3-142.  In  practice  it 
is 
often  difficult  to  predict  the 
optimum sweep rate. 
For 
this reason, the simplest approach 
is 
to 
start  with  the  optimum rate  set  using  the  ADJUST 
light.  Then,  while  observing  the  response  curve, gradually 
increase the sweep rate  until  the  amplitude or shape 
of 
the 
curve  begins  to  change.  When  the  curve  begins 
to 
change 
the sweep rate 
is 
too 
fast. 
3-143. 
External 
Triggering. 
3-144.  The  EXT TRIG IN connector enables the frequency 
sweep  to 
be 
remotely inhibited using  a contact closure  or 
TTL Logic  Levels.  This  signal  may  be  used to inhibit the 
sweep in the single, repetitive or Log Sweep Mode. 
3-145.  In  order to inhibit the sweep, the externally applied 
signal  into  the  EXT  TRIG  IN  connector 
is 
kept low.  To 
allow  the  3580A 
to 
perform  a  single  sweep,  the  inhibit 
signal 
is 
allowed  to 
go 
high for greater than 1 msec, 
but 
for 
less  than  the  total sweep  time. 
If 
the inhibit signal 
is 
not 
returned 
to 
low within the specified time, additional sweeps 
may be initiated. 
3-146.  To remotely inhibit the frequency  sweep  apply the 
following levels 
to 
the center terminal 
of 
the EXT TRIG IN 
connector: 
Sweep  Inhibit:  Ground 
(through< 
10 K)  or -0.5  V  de 
to 
0.5 V de. 
Sweep Enable:  Open or +2.5 V de 
to 
+5 
V de. 
NOTE 
The outer shield 
of 
the 
EXT 
TRIG 
IN 
connec-
tor 
is 
connected  to 
case 
ground.  The  center 
terminal 
of 
the  connector 
is 
the  inhibit 
line. 
3-19