Model 3580A 
BASIC 
OPERATING 
PROCEDURES 
Section III 
3-194.  Battery Operation (Option 001). 
a.  Connect  the  low {black) terminal 
of 
the front  panel 
INPUT  connector 
to 
earth  ground  or  to 
an 
appropriate 
system ground. 
b.  Set  the  POWER  switch  to  the 
ON 
{BAT)  position. 
The 
POWER 
light 
will 
illuminate. 
c.  Allow  approximately 
15 
seconds  for  the  CRT  to 
warm up. Adjust the  INTENSITY and FOCUS controls for 
a  bright,  clear  presentation  on  the  CRT.  When  the 
instrument 
is 
initially turned on, the display may 
be 
similar 
to  the  one  shown in  Figure  3-34. This  display  reflects the 
preferred  states 
of 
the  storage  elements  in  the  digital 
memory  and 
is 
not 
meaningful.  To  clear the  display,  press 
the CLEAR WRITE button. 
d.  Allow  a  warm-up  period 
of 
at  least  1 hour  before 
using the 3580A in a critical measurement application. 
e. 
To· 
recharge  the batteries, perform Steps a through d 
of 
the  power-line  turn on procedure {Paragraph 3-193 ). Set 
the 
POWER 
switch 
to 
the CHARGE position. The POWER 
light  will  illuminate. The  instrument cannot 
be 
used  while 
the batteries are being charged. 
The  instrument  should 
not 
be 
left 
in 
the 
CHARGE  mode  for  prolonged  periods.  A 
charge 
period 
of 
14  hours 
is 
sufficient  to 
recharge 
a  fully  discharged  battery  pack.  Ex-
tended periods 
of 
overcharge 
in 
ambient  tem-
peratures  exceeding 30" C ( 86° F) will severely 
degrade 
battery life and capacity by causing the 
cells to overheat. 
3-195. 
Frequency 
Calibration 
Procedure. 
3-196.  The  Frequency  Calibration  Procedure  should 
be 
performed  after  warm-up  each  time  the  instrument 
is 
turned  on. 
It 
should also 
be 
performed  before  and  after 
using the log sweep. 
3-197.  For operation in  the  Repetitive,  Single  or  Manual 
sweep mode, proceed 
as 
follows: 
a.  Turn  the  instrument  on 
as 
outlined  in  Paragraph 
3-192. 
b.  Set the 3580A controls 
as 
follows: 
ADAPTIVE 
SWEEP 
................. 
OFF 
DISPLAY 
..... 
STORE and BLANK STORE 
Released 
AMPLITUD~MODE 
....... 
LOG 10 dB/DIV 
AMPLITUDE REF LEVEL 
....... 
NORMAL 
dBv/LIN - dBm Switch 
........... 
dBv/LIN 
INPUT SENSITIVITY 
............... 
CAL 
VERNIER 
...................... 
CAL 
{Fully 
CW) 
FREQUENCY 
................. 
00.0 kHz 
START CTR 
.................... 
START 
BANDWIDTH 
.................... 
30 
Hz 
DISPLAY SMOOTHING 
............. 
MIN 
FREQ. 
SPAN/DN 
.................. 
N/A 
SWEEP 
TIME/DIV 
................. 
N/ A 
SWEEP 
MODE 
................... 
RESET 
c.  ·Clear  the  display  by  pressing  the  CLEAR  WRITE 
button. 
d.  Adjust the front panel ZERO 
CAL 
potentiometer for 
peak 
zero  response. (The zero response will  appear 
on 
the 
first line on the left-hand side 
of 
the display graticule.) 
e.  Set the BANDWIDTH to 10 Hz. Repeat Step 
d. 
3-198.  For operation in  the  Log  sweep mode, proceed 
as 
follows: 
a. 
Set the 3580A controls 
as 
follows: 
ADAPTIVE 
SWEEP 
................. 
OFF 
DISPLAY 
..... 
STORE and BLANK STORE 
Released 
AMPLITUDE MODE 
....... 
LOG 
10 dB/DIV 
AMPLITUDE REF LEVEL 
....... 
NORMAL 
dBv/LIN - dBm Switch 
........... 
dBv/LIN 
INPUT SENSITIVITY 
............... 
CAL 
VERNIER 
...................... 
CAL 
{Pully 
CW) 
FREQUENCY 
..................... 
N/ A 
START CTR 
...................... 
N/A 
BANDWIDTH 
. . . . . . . . . . . . . . . . . . . 
30 
Hz 
DISPLAY SMOOTHING 
............. 
MIN 
FREQ. SPAN/DIV 
.................. 
N/A 
SWEEP 
TIME/DIV 
................. 
N/ A 
SWEEP 
MODE 
............... 
LOG ZERO 
b.  Clear  the  display  by  pressing  the  CLEAR  WRITE 
button. 
c.  Adjust the front panel ZERO CAL potentiometer for 
peak zero  response. (The zero  response will  appear 
on 
the 
first line on the left-hand side 
of 
the display graticule.) 
3-199. 
Amplitude 
Calibration 
Procedure. 
3-200.  The  Amplitude  Calibration  Procedure  should 
be 
performed  initially  after  warm-up  and  each  time  the 
BANDWIDTH 
setting 
is 
changed. 
3-201.  For operation  on  the  1 
Hz 
or 3 Hz BANDWIDTH, 
proceed 
as 
follows: 
a. 
Tum  the  instrument  on  {Paragraph  3-192)  and 
perform  the  Frequency  Calibration  Procedure  (Paragraph 
3-195). 
3-27