Model 
3580A 
GENERAL 
OPERATING 
INFORMATION 
Section III 
~-~1IVERNIERI 
INPUT 
SENSITIVITY  Q 
q 
__ 
\ 
INPUT 
CIRCUITS 
IOOKHz  to  150KHz 
FROM 
VTO 
OVERLOAD 
DETECTOR 
I BANDWIDTH! 
MIXER 
Q 
I 
I 
I 
IF 
FILTER 
(IOOKHz) 
AMPLITUDE 
REF 
LEVEL 
0 
r----------~-----, 
I  I 
' 
LINEAR 
IF 
IF 
I 
AMPLIFIER  ATTENUATOR  I 
~--~1 
L----------------J 
VIDEO 
DETECTOR 
VIDEO  OUTPUT 
3580-
B-
3593 
VIDEO 
TO 
DISPLAY 
Figure 3-10.  Amplitude Section (Linear Mode). 
3-53.  The 
INPUT 
SENSITIVITY  switch  has 
12 
posi-
tions:  a 
CAL 
position  and 
11 
voltage  range  settings. 
With 
the  amplitude  VERNIER  in  the 
CAL 
position  and  the 
AMPLITUDE 
REF  LEVEL  switch  in the 
NORMAL 
(Xl) 
position,  the  full-scale  sensitivity, 
as 
determined  by  the 
INPUT 
SENSITIVITY switch setting, 
ranges 
from 20 V rms 
to 
0.2 m V 
rms. 
3-54.  For  any 
given 
INPUT  SENSITIVITY  setting,  the 
dynamic 
range 
of 
the  Input Circuits, 
Mixer 
and IF Filter 
is 
at least 80 
dB 
as 
long 
as 
the noise floor 
is 
more than 80 
dB 
below 
full 
scale.  Thus,  with  the 
INPUT 
SENSITIVITY 
switch 
in 
the  0.2 mV  position, 
an 
input 
signal 
as 
low 
as 
0.1 
µV 
rms could be detected at the output 
of 
the IF Filter. 
In  the  Linear  mode,  however,  the  dynamic 
range 
of 
the 
display 
is 
limited to approximately 40 
dB. 
This means that 
on  the  0.2 m V  range  the  smallest 
signal 
that  can  be 
displayed 
is 
approximately  2 µ V  or  1 % 
of 
full 
scale. 
Moreover,  the  2 µ V signal  might 
be 
visible  on the display 
but 
it 
would 
be 
too small to 
be 
measured accurately. For 
all 
practical  purposes,  then,  the  dynamic  display  range  is 
limited to approximately 20 
dB. 
3-55. 
To 
utilize  the 
full 
measurement 
range 
of 
the instru-
ment  in  the  Linear  mode,  it 
is 
necessary  to  increase  the 
display  sensitivity. 
To 
accomplish this a variable IF Attenu-
ator, controlled by the 
AMPLITUDE 
REF LEVEL switch, 
is 
inserted  between  the  Linear  IF  Amplifier  and  Video 
Detector. 
With 
the 
AMPLITUDE 
REF LEVEL switch set 
to 
the 
NORMAL 
(Xl) 
position,  the  IF attenuation 
is 
maxi-
mum. 
As 
the 
AMPLITUDE 
REF LEVEL switch 
is 
rotated 
in 
a clockwise  direction, the IF attenuation decreases,  the 
effective  IF 
gain 
increases  and  the  display  sensitivity 
increases.  The  IF  Attenuator  provides 
seven 
additional 
ranges 
which  allow  the  full-scale  sensitivity 
to 
be 
varied 
from 
0.1 
mV rms 
to 
0.1 
µV 
rms. 
3-56. 
By 
observing the 
INPUT 
SENSITIVITY and 
AMPLI-
TUDE 
REF  LEVEL  controls,  it  can  be  noted  that  the 
full-scale  (blue)  markings  on  the 
INPUT 
SENSITIVITY 
switch  dial  are  indfoated  by  a  white  window  that 
is 
mechanically  linked 
to 
the 
AMPLITUDE 
REF  LEVEL 
switch.  Changing the position 
of 
either switch changes the 
full-scale  sensitivity  in  a  20 
V, 
10 
V, 
2 
V, 
1 V sequence. 
Changing 
the 
AMPLITUDE 
REF LEVEL setting, however, 
does 
not change  the maximum  input  level.  For  example, 
with the 
INPUT 
SENSITIVITY switch set  for  a maximum 
input 
of 
1 V rms and the 
AMPLITUDE 
REF LEVEL switch 
set  to the 
X0.1 
position,  the  full-scale  sensitivity 
is 
0.1 
V 
rms,  the  display  sensitivity 
is 
10 
m V per  division 
but 
the 
maximum  input level 
is 
still 1 V rms.  Input signals greater 
than 
0.1 
V rms  but 
less 
than or equal  to 1 V rms will  not 
overdrive the mixer or produce an OVERLOAD indication. 
They  will,  however,  peak the display when the analyzer is 
tuned to their specific frequency. This does not damage the 
instrument  or hinder  its ability  to measure  signals  within 
the display range. 
3-57.  Using  the 
AMPLITUDE 
REF. 
LEVEL 
Control. 
Whenever 
possible,  the 
AMPLITUDE 
REF LEVEL switch 
should  be  left  in  the 
NORMAL 
(Xl) 
position  and  the 
INPUT 
SENSITIVITY  switch  should  be  used  to  set  the 
full-scale  sensitivity.  This 
is 
because the Amplitude Calibra-
tion  Procedure  (Paragraph  3-199) 
is 
performed  with the 
AMPLITUDE 
REF  LEVEL switch  in  the 
NORMAL 
posi-
tion  and  any  error  introduced  by  the  IF  Attenuator  is 
adjusted out. 
When 
the 
AMPLITUDE 
REF  LEVEL setting 
is 
changed from the 
NORMAL 
position, the accuracy 
of 
the 
IF  Attenuator  must  be  considered. 
This 
means  that  a 
possible  worst-case  error 
of 
± 
3% 
of 
full 
scale  must  be 
added  to the  amplitude accuracy specification.  Amplitude 
accuracy 
is 
discussed in Paragraph 3-72. 
3-58.  There  are  commonly  two  occasions  when  it 
is 
necessary  to change the 
AMPLITUDE 
REF LEVEL setting: 
a. 
When 
the  required  full-scale  sensitivity is  within the 
range 
of 
0.1 
mV rms  to  0.1 
µV 
rms and the amplitude 
of 
the input 
signal 
is 
less than or equal to 
0.1 
m V rms. In this 
case, the 
INPUT 
SENSITIVITY switch 
is 
set to the 0.2 m V 
range  (fully clockwise) and the appropriate range is selected 
using the 
AMPLITUDE 
REF LEVEL switch. 
b.  For  expanded-scale  measurements  where  the ampli· 
tude 
of 
the  input 
signal 
is  0.2 m V rms or greater and the 
signal 
or 
signals 
of 
interest are 
less 
than 
10% 
of 
full  scale 
with  the 
INPUT 
SENSITIVITY  switch  set  to  the  lowest 
range  that does  not produce an OVERLOAD indication. 
In 
this case, the 
AMPLITUDE 
REF LEVEL switch is initially 
set  to  the 
Xl 
position  and  the 
INPUT 
SENSITIVITY 
switch  is set  to the lowest range  that does  not produce an 
3-9