Model 3580A 
GENERAL 
OPERATING 
INFORMATION 
Section III 
0 Hz  1 kHz 
Figure 
3-20.  Zero 
Response 
(300 
Hz 
BW). 
3-96. 
Frequency 
Setting. 
3-97.  The  front  panel  FREQUENCY  control  tunes  the 
frequency 
of 
the  analyzer  over the  0 
Hz 
to 
50 kHz range. 
The  control can  be  used 
to 
set  either the start 
or 
center 
frequency 
of 
a linear sweep.  The  start or center frequency 
selected  by  the  FREQUENCY  control 
is 
indicated on the 
FREQUENCY dial. 
3-98.  The  FREQUENCY  control has  two  selectable  drive 
ratios 
to 
permit  coarse  or  fine  tuning.  Coarse  tuning 
is 
selected by pushing the  crank toward the front  panel; fine 
tuning 
is 
selected  by  pulling  the  crank outward.  In  the 
coarse  position,  one  revolution 
of 
the  crank  changes  the 
FREQUENCY  dial  setting  by  approximately  2.7 kHz.  In 
the  fine  position, one  revolution 
of 
the crank changes the 
frequency by approximately 73 Hz. 
3-99. 
Frequency 
Dial.  The  FREQUENCY  dial  indicates 
the  start  or  center frequency  in  kHz.  Dial  settings  range 
from  00.0 kHz 
to 
approximately  50.5 kHz.  The  frequency 
dial  resolution  is  20 Hz  represented by  one minor  division 
on  the  frequency  scale.  When  the  instrument 
is 
properly 
calibrated (Paragraph  3-195),  the  frequency  dial  accuracy 
is: 
a.  ±  100 Hz  when  the  ambient temperature 
is 
within 
the  range 
of 
20" C  ( 68° 
F) 
to 
30° C 
(86°F). 
b.  ± 
300 
Hz  when the  ambient temperature is within 
the  range 
of 
0° C  (32° 
F) 
to 
20° C 
(68° 
F) 
or 
30" C  (86° 
F) 
to 
55° C 
(131° F). 
3-100.  Start/Center.  With  the  START/CENTER  slide 
switch  in  the  START  position,  the  FREQUENCY  dial 
setting  indicates  the  frequency  represented  by  the  first 
vertical  line  on  the  left-hand side 
of 
the  display graticule. 
This 
is 
the  "start  frequency"  or frequency  at which  the 
sweep begins. With the switch in the CENTER position, the 
FREQUENCY  dial  setting  indicates  the  frequency  repre-
sented by  the  center vertical line 
on 
the display graticule. 
This is the "center frequency" 
of 
the sweep. 
3-101. 
When 
surveying a spectrum containing two 
or 
more 
signals,  it 
is 
generally  convenient 
to 
leave  the  START/ 
CENTER  switch  in  the  START  position.  The  FRE-
QUENCY  control  can  then  be  used 
to 
set  the  start 
frequency  and  the  FREQUENCY 
SP 
AN 
control  can  be 
used 
to 
set  the  spectrum  width  or  "end frequency".  To 
observe  one  frequency  component in  a spectrum, set  the 
START/CENTER switch 
to 
the CENTER position and set 
the  FREQUENCY  dial 
to 
the  frequency 
of 
interest.  The 
frequency 
of 
interest  will  appear  in  the  center 
of 
the 
display.  The  width 
of 
the center frequency response can be 
adjusted  by  changing  the  FREQUENCY 
SP 
AN 
or BAND-
WIDTH 
setting. 
3-102. 
Zero 
Cal. 
Potentiometer.  The  purpose 
of 
the 
ZERO  CAL  potentiometer 
is 
to 
enable  the  operator  to 
compensate· for slight variations in frequency  dial accuracy 
that 
occur  during  warm-up  or  when 
the 
instrument 
is 
operated 
in 
an uncontrolled environment. The  ZERO CAL 
potentiometer 
is 
also used in the Log Zero sweep mode 
to 
establish the correct starting point for the log sweep. Refer 
to 
Paragraph  3-195  for  the  Frequency  Calibration  Proce-
dure. 
3-103. 
Frequency 
Span 
Setting. 
3-104.  The  FREQUENCY SPAN  control sets the  width 
of 
the  spectrum  to  be  observed  during  linear  or  manual 
sweeps.  Excluding  the  0 
Hz 
position,  there  are  ten  FRE-
QUENCY 
SP 
AN 
settings ranging from  5 Hz  per  division 
to 
5 kHz  per  division.  With  ten  horizontal  divisions  on  the 
display,  the  overall  spectrum  width  can  be  adjusted from 
50 
Hz 
to 
50 kHz. 
3-105.  0 
Hz 
Span. 
With  the  FREQUENCY  SPAN switch 
set 
to 
the 0 Hz position, the instrument remains at the start 
or  center  frequency  indicated  on  the  FREQUENCY  dial. 
The  display,  however,  continues 
to 
sweep  at  the  rate 
selected 
by 
the 
SWEEP 
TIME  setting.  The  result  is  a 
graphical display 
of 
amplitude 
vs. 
time. 
3-106.  The  amplitude vs.  time feature 
is 
useful for observ-
ing  the  amplitude  variations 
of 
a  signal 
that 
occur  over 
relatively long periods 
of 
time. For example, the amplitude 
of 
the  10 kHz  sine 
wave 
shown  in  Figure  3-21 A  appears 
stable on a conventional oscilloscope 
but 
is actually varying 
at a very slow rate. In Figure 3-21B, the 3580A 
was 
used 
to 
monitor the  amplitude 
of 
the  10 kHz  signal  over a 2,000 
second period. The 3580A amplitude 
vs. 
time display shows 
that  the  10 kHz  signal 
is 
amplitude  modulated  by  a 
triangular-shaped signal whose frequency 
is 
0.00166 Hz. 
3-107.  Because 
of 
its narrow bandwidth, the 3580A cannot 
respond 
to 
rapid  changes  in  amplitude.  The  maximum 
modulating frequency  that can  be observed and measured 
with any accuracy 
is 
approximately  100 Hz  on the 
300 
Hz 
BANDWIDTH 
setting. 
3-15