EasyManua.ls Logo

Intersoft Electronics RFA641 - Figure 76: Adjust Trigger Level (RFM Setup)

Intersoft Electronics RFA641
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Radar Field Analyser - RFA641 Edition Date: 28-Feb-18
A number of indicators display the detected interrogations:
The level indicators show the instantaneous levels of the interrogation pulses as they are identified
as P1, P2 or P3. For P3 the levels are indicated depending on the detected interrogation timing.
Trigger Count: Each valid pulse is indicated in the trigger count.
Reply Count: Incremented each time a reply is sent
P2 Count: Single pulses and pulses that correspond to the 2 microsecond delay for a P2 pulse are
counted
Mode 1,2,3/A,C: Counter for the specific interrogation modes detected (and replies sent).
To stop the RFM operation, click the Stop button. In case you want the RFM to keep on running but
want to disconnect the computer, you can simply click the Done button. The software will stop without
however stopping the device’s RFM operation.
The RFM can also operate in a moving target mode where it will fly a simulated target in range from the
radar. This is invoked by selecting Moving Target in the Mode selection on the Remote Field Monitor.vi.
When selected the Define Speed Scenario window will be invoked.
IE-UM-00165-005 RFA641 User Manual.odt 70/110
Figure 76: Adjust trigger level (RFM Setup)

Table of Contents