MULTICAL®603
10
KamstrupA/S∙Technicaldescription∙55122029_A1_GB_06.2017
2 Technicaldata
2.1 Approvedmeterdata
Approvals DK0200MI004040,heatmeter
accordingtoMID2014/32EUandEN1434:2015
TS27.02012,coolingmeterandcooling/heatmeter
AccordingtoDKBEK1178andEN1434:2015
EUdirectives MeasuringInstrumentsDirective,LowVoltageDirective,
ElectromagneticCompatibilityDirective,PressurizedEquipmentDirective,
RadioEquipmentDirective
Heatmeterapproval DK0200MI004040
Thestatedminimumtemperaturesarerelatedto
thetypeapproval.
Themeterhasnocutoffforlowtemperature
andthusmeasuresdownto0.01°Cand0.01K.
Temperaturerange:2°C…180°C
Differentialrange :3K…178K
Coolingmeter
andcooling/heatmeter TS27.02012
Temperaturerange:2°C…180°C
Differentialrange :3K…178K
Mediatemperature,ULTRAFLOW®
q
:2°C…130°C
Accuracy
- Calculator E
c
=(0.5+
min
/)%
- Flowsensor,ULTRAFLOW® E
f
=(2+0.02qp/q),butnotexceeding±5%
Temperaturesensorconnection Type603A Pt100EN60751,2wireconnection
Type603B Pt100EN60751,4wireconnection
Type603C/E/F Pt500EN60751,2wireconnection
Type603
D/G Pt500EN60751,4wireconnection
EN1434designation EnvironmentalclassAandC
MIDdesignation Mechanicalenvironment:ClassM1andM2
Electromagneticenvironment:ClassE1andE2
Noncondensingenvironment,closedlocation
(indoors),5…55°C

Table of Contents

Other manuals for Kamstrup MULTICAL 603

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kamstrup MULTICAL 603 and is the answer not in the manual?

Kamstrup MULTICAL 603 Specifications

General IconGeneral
Flow SensorUltrasonic
Battery LifeUp to 16 years
Power SupplyBattery
Communication OptionsM-Bus, Wireless M-Bus, Modbus, BACnet
DisplayLCD
ApplicationHeating and Cooling
Pressure RatingPN16
ApprovalMID
AccuracyClass 2
Ambient Temperature Range55°C
Housing MaterialBrass

Related product manuals