2 Technicaldata
2.1 Approvedmeterdata
Approvals DK‐0200‐MI004‐040,heatmeter
accordingtoMID2014/32EUandEN1434:2015
TS27.02012,coolingmeterandcooling/heatmeter
AccordingtoDK‐BEK1178andEN1434:2015
EUdirectives MeasuringInstrumentsDirective,LowVoltageDirective,
ElectromagneticCompatibilityDirective,PressurizedEquipmentDirective,
RadioEquipmentDirective
Heatmeterapproval DK‐0200‐MI004‐040
Thestatedminimumtemperaturesarerelatedto
thetypeapproval.
Themeterhasnocut‐offforlowtemperature
andthusmeasuresdownto0.01°Cand0.01K.
Temperaturerange:2°C…180°C
Differentialrange :3K…178K
Coolingmeter
andcooling/heatmeter TS27.02012
Temperaturerange:2°C…180°C
Differentialrange :3K…178K
Mediatemperature,ULTRAFLOW®
q
:2°C…130°C
Accuracy
- Calculator E
c
=(0.5+
min
/)%
- Flowsensor,ULTRAFLOW® E
f
=(2+0.02qp/q),butnotexceeding±5%
Temperaturesensorconnection Type603‐A Pt100–EN60751,2‐wireconnection
Type603‐B Pt100–EN60751,4‐wireconnection
Type603‐C/E/F Pt500–EN60751,2‐wireconnection
Type603
‐D/G Pt500–EN60751,4‐wireconnection
EN1434designation EnvironmentalclassAandC
MIDdesignation Mechanicalenvironment:ClassM1andM2
Electromagneticenvironment:ClassE1andE2
Non‐condensingenvironment,closedlocation
(indoors),5…55°C