Trueenergycalculation
Duringtestandverification,theheatmeter’senergycalculationiscomparedtothe“trueenergy”,whichiscalculated
accordingtotheformulaofEN1434‐1:2007,EN1434‐1:2015andOIMLR75:2002.
Thebelow‐mentionedenergycalculatorcanbesuppliedelectronicallybyKamstrupA/S.
Thetrueenergyatthe
mostfrequentlyusedverificationpointsisindicatedinthetablebelow:
t1C t2C K
Inlet
Wh/0.1m
3
Outlet
Wh/0.1m
3
42 40 2230.11 230.29
43 40 3345.02 345.43
53 50 3343.62 344.11
50 40 101146.70 1151.55
70 50 202272.03 2295.86
80 60 202261.08 2287.57
160 40 12012793.12 13988.44
160 20 14014900.00 16390.83