Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-113
Usage
f, msg = InitiatePulseTest(tag)
f
A Boolean flag; this flag is true when the pulse was successfully configured,
false
when errors are encountered
msg
A string message; if the f flag is false, msg contains an error message; if it is
true, msg contains a string that indicates successful configuration
tag
Numeric identifier of the pulse configuration to be initiated
Details
This function only initiates configured pulse trains assigned to a valid tag. Configure the pulse before
initiating it using one of the ConfigurePulse* functions (refer to the Also see section).
Example
smua.reset()
smua.source.rangev = 5
smua.source.rangei = 1
smua.source.levelv = 0
smua.measure.rangev = 5
smua.measure.rangei = 1
smua.measure.nplc = 0.01
smua.measure.autozero = smua.AUTOZERO_ONCE
smua.nvbuffer1.clear()
smua.nvbuffer1.appendmode = 1
smua.source.output = smua.OUTPUT_ON
f1, msg1 = ConfigPulseVMeasureI(smua, 0, 5,
1, 0.002, 0.2, 10, smua.nvbuffer1, 1)
if f1 == true then
f2, msg2 = InitiatePulseTest(1)
print("Initiate message:", msg2)
else
print("Config errors:", msg1)
end
Configure System SourceMeter
®
instrument channel A to generate
a pulse train. If no errors are
encountered, initiate the pulse
train. Channel A pulses voltage
from a bias level of 0 V to a pulse
level of 5 V. The pulse level is
present for 2 ms and the bias level
for 200 ms, with a 1 A limit setting.
A total of 10 pulses is generated,
and the measurement data is
stored in smua.nvbuffer1. This
pulse train is assigned to tag =
1.
Also see
ConfigPulseIMeasureV() (on page 7-36)
ConfigPulseVMeasureI() (on page 7-42)
ConfigPulseIMeasureVSweepLin() (on page 7-38)
ConfigPulseVMeasureISweepLin() (on page 7-45)
ConfigPulseIMeasureVSweepLog() (on page 7-40)
ConfigPulseVMeasureISweepLog() (on page 7-47)
KIPulse factory script (on page 5-22)