7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-378 2600BS-901-01 Rev. C / August 2016
Details
When pass-through mode is enabled, triggers are passed through immediately and initiate the delay.
When disabled, a trigger only initiates a delay.
Example
trigger.timer[1].passthrough = true
Enables pass-through mode on trigger timer 1.
Also see
trigger.timer[N].reset() (on page 7-378)
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.timer[N].reset()
N
Trigger timer number (1 to 8)
Details
The trigger.timer[N].reset() function resets the following attributes to their factory defaults:
trigger.timer[N].count
trigger.timer[N].delay
trigger.timer[N].delaylist
trigger.timer[N].passthrough
trigger.timer[N].stimulus
It also clears trigger.timer[N].overrun.
Example
trigger.timer[1].reset()
Resets the attributes associated with timer 1 back
to factory default values.
Also see
trigger.timer[N].count (on page 7-374)
trigger.timer[N].delay (on page 7-375)
trigger.timer[N].delaylist (on page 7-375)
trigger.timer[N].overrun (on page 7-377)
trigger.timer[N].passthrough (on page 7-377)
trigger.timer[N].stimulus (on page 7-379)