EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #717 background imageLoading...
Page #717 background image
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-378 2600BS-901-01 Rev. C / August 2016
Details
When pass-through mode is enabled, triggers are passed through immediately and initiate the delay.
When disabled, a trigger only initiates a delay.
Example
trigger.timer[1].passthrough = true
Enables pass-through mode on trigger timer 1.
Also see
trigger.timer[N].reset() (on page 7-378)
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.timer[N].reset()
N
Trigger timer number (1 to 8)
Details
The trigger.timer[N].reset() function resets the following attributes to their factory defaults:
• trigger.timer[N].count
• trigger.timer[N].delay
• trigger.timer[N].delaylist
• trigger.timer[N].passthrough
• trigger.timer[N].stimulus
It also clears trigger.timer[N].overrun.
Example
trigger.timer[1].reset()
Resets the attributes associated with timer 1 back
to factory default values.
Also see
trigger.timer[N].count (on page 7-374)
trigger.timer[N].delay (on page 7-375)
trigger.timer[N].delaylist (on page 7-375)
trigger.timer[N].overrun (on page 7-377)
trigger.timer[N].passthrough (on page 7-377)
trigger.timer[N].stimulus (on page 7-379)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600B and is the answer not in the manual?

Keithley 2600B Specifications

General IconGeneral
BrandKeithley
Model2600B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals