7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-294 2600BS-901-01 Rev. C / August 2016
Example 2
-- decimal 20480 = binary 0101 0000 0000 0000
operationRegister = 20480
status.operation.enable = operationRegister
Sets the USER and PROG bits of the
operation status enable register using a
decimal value.
Also see
Operation Status Registers (on page E-9)
status.operation.calibrating.*
This attribute contains the operation status calibration summary register set.
Type TSP-Link accessible Affected by Where saved Default value
Attribute - - - - - - - -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B: 6
(All bits set)
Usage
operationRegister = status.operation.calibrating.condition
operationRegister = status.operation.calibrating.enable
operationRegister = status.operation.calibrating.event
operationRegister = status.operation.calibrating.ntr
operationRegister = status.operation.calibrating.ptr
status.operation.calibrating.enable = operationRegister
status.operation.calibrating.ntr = operationRegister
status.operation.calibrating.ptr = operationRegister
operationRegister
The status of the operation calibrating event register; a zero (0) indicates no bits set
(also send 0 to clear all bits); other values indicate various bit settings