7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-344 2600BS-901-01 Rev. C / August 2016
status.reset()
This function resets all bits in the status model.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
status.reset()
Details
This function clears all status data structure registers (enable, event, NTR, and PTR) to their default
values. For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status
register set contents (on page E-1) and Enable and transition registers (on page E-19).
Example
status.reset()
Resets the instrument status model.
Also see
Status model (on page 5-15, on page E-1)
status.standard.*
These attributes manage the standard event status register set of the status model.
Type TSP-Link accessible Affected by Where saved Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R) Yes Status reset Not saved 0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
253 (All bits set)
Usage
standardRegister = status.standard.condition
standardRegister = status.standard.enable
standardRegister = status.standard.event
standardRegister = status.standard.ntr
standardRegister = status.standard.ptr
status.standard.enable = standardRegister
status.standard.ntr = standardRegister
status.standard.ptr = standardRegister
standardRegister
The status of the standard event status register; a zero (0) indicates no bits set (also
send 0 to clear all bits); other values indicate various bit settings