Details
These attributes are used to read or write to the standard event status registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are
set. The least significant bit of the binary number is bit B0, and the most significant bit is bit B15. For
example, if a value of 1.29000e+02 (which is 129) is read as the value of the condition register, the
binary equivalent is 0000 0000 1000 0001. This value indicates that bit B0 and bit B7 are set.
B15
B14
B13
B12
B11
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
>
>
>
>
>
>
>
>
*
0
0
0
0
0
0
0
0
1
0
0
0
0
0
0
1
* Least significant bit