Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-393
Details
Releases a trigger that was asserted with an indefinite pulse width. It also releases a trigger that was
latched in response to receiving a synchronous mode trigger.
Example
tsplink.trigger[3].release()
Releases trigger line 3.
Also see
tsplink.trigger[N].assert() (on page 7-387)
tsplink.trigger[N].clear() (on page 7-388)
tsplink.trigger[N].mode (on page 7-389)
tsplink.trigger[N].overrun (on page 7-391)
tsplink.trigger[N].pulsewidth (on page 7-392)
tsplink.trigger[N].stimulus (on page 7-394)
tsplink.trigger[N].wait() (on page 7-396)
tsplink.trigger[N].reset()
This function resets some of the TSP-Link trigger attributes to their factory defaults. This function is not available
on the Models 2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
tsplink.trigger[N].reset()
N
The trigger line (1 to 3)
Details
The tsplink.trigger[N].reset() function resets the following attributes to their factory
defaults:
tsplink.trigger[N].mode
tsplink.trigger[N].stimulus
tsplink.trigger[N].pulsewidth
This also clears tsplink.trigger[N].overrun.
Example
tsplink.trigger[3].reset()
Resets TSP-Link trigger line 3
attributes back to factory default
values.
Also see
tsplink.trigger[N].mode (on page 7-389)
tsplink.trigger[N].overrun (on page 7-391)
tsplink.trigger[N].pulsewidth (on page 7-392)
tsplink.trigger[N].stimulus (on page 7-394)