7: TSP command reference Series 2600B System SourceMeter® Instrument
7-220 2600BS-901-01 Rev. C / August 2016
Details
This attribute sets the size of the stack used for filtered measurements.
Example
smua.measure.filter.count = 10
Sets the filter count for SMU channel A to 10.
Also see
Filters (on page 3-3)
reset() (on page 7-173)
setup.recall() (on page 7-194)
smuX.measure.filter.enable (on page 7-220)
smuX.measure.filter.type (on page 7-221)
smuX.reset() (on page 7-232)
smuX.measure.filter.enable
This command enables or disables filtered measurements.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
SMU reset
Saved setup 0 (smuX.FILTER_OFF)
Usage
filterState = smuX.measure.filter.enable
smuX.measure.filter.enable = filterState
The filter status; set to one of the following values:
0 or smuX.FILTER_OFF: Disables the filter
X
.FILTER_ON: Enables the filter
SMU channel (for example,
smua.measure.filter.enable
applies to SMU
Details
This command enables or disables the filter.
Example
smua.measure.filter.enable = 1
Enables the filter for SMU channel A. Alternatively, the
value 1 may be replaced with
.
Also see
Filters (on page 3-3)
reset() (on page 7-173)
setup.recall() (on page 7-194)
smuX.measure.filter.count (on page 7-219)
smuX.measure.filter.type (on page 7-221)
smuX.reset() (on page 7-232)