Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-94 2600BS-901-01 Rev. C / August 2016
eventlog.count
This attribute returns the number of events in the event log.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Instrument reset
Clearing event log
Reading event log
Not applicable Not applicable
Usage
N = eventlog.count
N
The number of events in the event log
Example
print(eventlog.count)
Displays the present number of events in the
instrument event log.
Output looks similar to:
3.00000e+00
Also see
eventlog.all() (on page 7-92)
eventlog.clear() (on page 7-93)
eventlog.enable (on page 7-94)
eventlog.next() (on page 7-95)
eventlog.overwritemethod (on page 7-96)
eventlog.enable
This attribute enables or disables the event log.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Not saved
1 (eventlog.ENABLE)
Usage
status = eventlog.enable
eventlog.enable = status
status
The enable status of the event log:
1 or eventlog.ENABLE: Event log enable
0 or eventlog.DISABLE: Event log disable
Details
When the event log is disabled (eventlog.DISABLE or 0), no new events are added to the event
log. You can, however, read and remove existing events.
When the event log is enabled, new events are logged.