Page #515 background imageLoading...
Page #515 background image
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-176 2600BS-901-01 Rev. C / August 2016
script.delete()
This function deletes a script from nonvolatile memory.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
script.delete(scriptName)
scriptName
The string that represents the name of the script
Example
script.delete("test8")
Deletes a user script named "test8" from
nonvolatile memory.
Also see
Delete user scripts (on page 6-9)
Delete user scripts from the instrument (on page 6-44)
scriptVar.save() (on page 7-186)
script.factory.catalog()
This function returns an iterator that can be used in a for loop to iterate over all the factory scripts.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
for name in script.factory.catalog() do body end
name
String representing the name of the script
body
Code that implements the body of the for loop to process the names in the catalog
Details
Accessing this catalog of scripts allows you to process the factory scripts. The entries will be
enumerated in no particular order.
Each time the body of the function executes, name takes on the name of one of the factory scripts.
The for loop repeats until all scripts have been iterated.
Example
for name in script.factory.catalog() do
print(name)
end
Retrieve the catalog listing for factory scripts.
Also see
None