EasyManua.ls Logo

Keithley 2600B

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-397
Details
Use tsplink.writebit() and tsplink.writeport() to control the output state of the trigger
line when trigger operation is set to tsplink.TRIG_BYPASS.
If the output line is write-protected by the tsplink.writeprotect attribute, this command is
ignored.
The reset function does not affect the present states of the TSP-Link trigger lines.
Example
tsplink.writebit(3, 0)
Sets trigger line 3 low (0).
Also see
tsplink.readbit() (on page 7-384)
tsplink.readport() (on page 7-384)
tsplink.writeport() (on page 7-397)
tsplink.writeprotect (on page 7-398)
tsplink.writeport()
This function writes to all TSP-Link synchronization lines. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
tsplink.writeport(data)
data
Value to write to the port (0 to 7)

Table of Contents

Related product manuals