Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-397
Details
Use tsplink.writebit() and tsplink.writeport() to control the output state of the trigger
line when trigger operation is set to tsplink.TRIG_BYPASS.
If the output line is write-protected by the tsplink.writeprotect attribute, this command is
ignored.
The reset function does not affect the present states of the TSP-Link trigger lines.
Example
tsplink.writebit(3, 0)
Sets trigger line 3 low (0).
Also see
tsplink.readbit() (on page 7-384)
tsplink.readport() (on page 7-384)
tsplink.writeport() (on page 7-397)
tsplink.writeprotect (on page 7-398)
tsplink.writeport()
This function writes to all TSP-Link synchronization lines. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
tsplink.writeport(data)
data
Value to write to the port (0 to 7)