Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-87
display.trigger.overrun
This attribute contains the event detector overrun status.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Display trigger clear
Instrument reset
Recall setup
Not saved
false
Usage
overrun = display.trigger.overrun
overrun
The trigger overrun state
Details
Indicates if a trigger event was ignored because the event detector was already in the detected state
when the TRIG button was pressed.
Indicates the overrun state of the event detector built into the display.
This attribute does not indicate whether an overrun occurred in any other part of the trigger model or
in any other detector that is monitoring the event.
Example
overrun = display.trigger.overrun
Sets the variable overrun equal to the
present state of the event detector built into
the display.
Also see
display.trigger.clear() (on page 7-86)
display.trigger.wait()
This function waits for the TRIG key on the front panel to be pressed.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
triggered = display.trigger.wait(timeout)
triggered
true: Trigger was detected
false: The operation timed out
timeout
Timeout in seconds