Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-389
Details
This number is used by the TSP-Link trigger line when it detects an input trigger.
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
trigger events from this line.
Example
trigger.timer[1].stimulus = tsplink.trigger[2].EVENT_ID
Sets the trigger stimulus
of trigger timer 1 to the
TSP-Link trigger 2 event.
Also see
None
tsplink.trigger[N].mode
This attribute defines the trigger operation and detection mode. This attribute is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
TSP-Link trigger N reset
Not saved
0 (tsplink.TRIG_BYPASS)
Usage
mode = tsplink.trigger[N].mode
tsplink.trigger[N].mode = mode
mode
The trigger mode
N
The trigger line (1 to 3)
Details
This attribute controls the mode in which the trigger event detector and the output trigger generator
operate on the given trigger line.