System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. C / August 2016 7-217
Example
smua.measure.calibratev(1, 1e-4, 1e-5, 0.92,
0.903)
SMU channel A calibrates voltage
measurement using following values: 1 V
calibration range, 1e−4 for +zero
measurement reading, 1e−5 for +zero
DMM measurement reading, 0.92 for
+FS measurement read
the +FS DMM measurement reading.
Also see
Adjustment (on page B-18)
smuX.cal.lock() (on page 7-201)
smuX.cal.restore() (on page 7-203)
smuX.cal.save() (on page 7-204)
smuX.cal.unlock() (on page 7-205)
smuX.source.calibrateY() (on page 7-235)
smuX.measure.count
This attribute sets the number of measurements performed when a measurement is requested.
Type TSP-Link accessible Affected by Where saved Default value
SMU reset
Usage
count = smuX.measure.count
smuX.measure.count = count
Source-measure unit (SMU) channel (for example,
applies
Details
This attribute controls the number of measurements taken any time a measurement is requested.
When using a reading buffer with a measure command, this attribute also controls the number of
readings to be stored.
If smuX.measure.count is set to a value greater than 1, any measurement delay set by
smuX.measure.delay will only occur before the first measurement, while the
smuX.measure.interval controls the interval between successive measurements.
Example
Sets the SMU channel A measure count to 10.
Also see
reset() (on page 7-173)
setup.recall() (on page 7-194)
smuX.measure.delay (on page 7-218)
smuX.measure.interval (on page 7-222)
smuX.measure.overlappedY() (on page 7-225)
smuX.measure.Y() (on page 7-229)
smuX.reset() (on page 7-232)