EasyManua.ls Logo

Keithley 2600B

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-358 2600BS-901-01 Rev. C / August 2016
Performs a linear current sweep with voltage measured at every step (point):
1. Sets the smu to output starti amperes, allows the source to settle for stime seconds, and then
performs a voltage measurement.
2. Sets the smu to output the next amperes step, allows the source to settle for stime seconds, and then
performs a voltage measurement.
3. Repeats the above sequence until the voltage is measured on the stopi amperes step.
The linear step size is automatically calculated as follows:
step = (stopi starti) / (points 1)
Figure 134: SweepILinMeasureV()
Example
SweepILinMeasureV(smua, -1E-3, 1E-3, 0, 100)
This function performs a 100-point linear current
sweep starting at 1 mA and stopping at +
1 mA.
Voltage is measured at every step (point) in the
sweep. Because stime is set for 0 s, voltage is
measured as quickly as possible after each
current step.
Also see
KISweep factory script (on page 5-21)
SweepIListMeasureV()
This KISweep factory script function performs a current list sweep with voltage measured at every step (point).
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
SweepIListMeasureV(smu, ilist, stime, points)
smu
System SourceMeter
®
instrument channel (for example, smua refers to SMU
channel A)
ilist
Arbitrary list of current source values; ilist = {value1, value2,
...valueN}
stime
Settling time in seconds; occurs after stepping the source and before performing a
measurement
points
Number of sweep points (must be
2)

Table of Contents

Related product manuals