EasyManuals Logo

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #730 background imageLoading...
Page #730 background image
Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-391
Also see
digio.writebit() (on page 7-64)
digio.writeport() (on page 7-64)
tsplink.trigger[N].assert() (on page 7-387)
tsplink.trigger[N].clear() (on page 7-388)
tsplink.trigger[N].overrun (on page 7-391)
tsplink.trigger[N].release() (on page 7-392)
tsplink.trigger[N].reset() (on page 7-393)
tsplink.trigger[N].stimulus (on page 7-394)
tsplink.trigger[N].wait() (on page 7-396)
tsplink.trigger[N].overrun
This attribute indicates if the event detector ignored an event while in the detected state. This attribute is not
available on the Models 2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Instrument reset
Recall setup
TSP-Link trigger N clear
TSP-Link trigger N reset
Not applicable
Not applicable
Usage
overrun = tsplink.trigger[N].overrun
overrun
Trigger overrun state
N
The trigger line (1 to 3)
Details
Indicates that an event was ignored because the event detector was in the detected state when the
event was detected.
Indicates the overrun state of the event detector built into the line itself.
It does not indicate whether an overrun occurred in any other part of the trigger model or in any other
detector that is monitoring the event.
It does not indicate output trigger overrun. Output trigger overrun indications are provided in the
status model.
Example
print(tsplink.trigger[1].overrun)
If an event was ignored, displays
true; if an event was not ignored,
displays
false
.
Also see
tsplink.trigger[N].assert() (on page 7-387)
tsplink.trigger[N].clear() (on page 7-388)
tsplink.trigger[N].mode (on page 7-389)
tsplink.trigger[N].release() (on page 7-392)
tsplink.trigger[N].reset() (on page 7-393)
tsplink.trigger[N].stimulus (on page 7-394)
tsplink.trigger[N].wait() (on page 7-396)