System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. C / August 2016 7-283
Bit Value Description
status.measurement.instrument.SMUA
Set bit indicates one or more enabled bits of
the measurement event SMU A summary
register is set.
Bit B1 decimal value: 2
B2
status.measurement.instrument.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set
bit indicates one or more enabled bits of the
measurement event SMU B summary register
is set.
Bit B2 decimal value: 4
As an example, to set bit B1 of the measurement event instrument summary enable register, set
status.measurement.instrument.enable = status.measurement.instrument.SMUA.
In addition to the above constants, measurementRegister can be set to the decimal equivalent of
the bit to set. To set more than one bit of the register, set measurementRegister to the sum of
their decimal weights. For example, to set bits B1 and B2, set measurementRegister to 6 (which is
the sum of 2 + 4).
Example
status.measurement.instrument.enable =
status.measurement.instrument.SMUA
Sets the SMU A bit of the
measurement event instrument
summary enable register using a
Also see
Measurement event registers (on page E-8)
status.measurement.instrument.smuX.*
This attribute contains the registers of the measurement event SMU X summary register set.
Type TSP-Link accessible Affected by Where saved Default value
Attribute - - - - - - - -
.enable (RW) Yes Status reset Not saved 0
.ptr (RW) Yes Status reset Not saved 387 (All bits set)