7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-60 2600BS-901-01 Rev. C / August 2016
digio.trigger[N].reset()
This function resets trigger values to their factory defaults. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
digio.trigger[N].reset()
N
Digital I/O trigger line (1 to 14)
Details
This function resets the following attributes to factory default settings:
digio.trigger[N].mode
digio.trigger[N].pulsewidth
digio.trigger[N].stimulus
It also clears digio.trigger[N].overrun.
Example
digio.trigger[3].mode = 2
digio.trigger[3].pulsewidth = 50e-6
digio.trigger[3].stimulus = digio.trigger[5].EVENT_ID
print(digio.trigger[3].mode, digio.trigger[3].pulsewidth,
digio.trigger[3].stimulus)
digio.trigger[3].reset()
print(digio.trigger[3].mode, digio.trigger[3].pulsewidth,
digio.trigger[3].stimulus)
Set the digital I/O trigger line 3 for a falling edge with a pulsewidth of 50 microseconds.
Use digital I/O line 5 to trigger the event on line 3.
Reset the line back to factory default values.
Output before reset:
2.00000e+00 5.00000e-05 5.00000e+00
Output after reset:
0.00000e+00 1.00000e-05 0.00000e+00
Also see
digio.trigger[N].mode (on page 7-56)
digio.trigger[N].overrun (on page 7-58)
digio.trigger[N].pulsewidth (on page 7-58)
digio.trigger[N].stimulus (on page 7-61)