Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-384 2600BS-901-01 Rev. C / August 2016
Also see
tsplink.reset() (on page 7-385) tsplink.reset() (on page 7-385)
tsplink.state (on page 7-386)
tsplink.readbit()
This function reads the state of a TSP-Link synchronization line. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
data = tsplink.readbit(N)
data
The state of the synchronization line
N
The trigger line (1 to 3)
Details
Returns a value of zero (0) if the line is low and 1 if the line is high.
Example
data = tsplink.readbit(3)
print(data)
Assume line 3 is set high, and it is then read.
Output:
1.000000e+00
Also see
tsplink.readport() (on page 7-384)
tsplink.writebit() (on page 7-396)
tsplink.readport()
This function reads the TSP-Link synchronization lines as a digital I/O port. This function is not available on the
Models 2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
data = tsplink.readport()
data
Numeric value that indicates which lines are set
Details
The binary equivalent of the returned value indicates the input pattern on the I/O port. The least
significant bit of the binary number corresponds to line 1 and the value of bit 3 corresponds to line 3.
For example, a returned value of 2 has a binary equivalent of 010. This indicates that line 2 is high
(1), and that the other two lines are low (0).