Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-404 2600BS-901-01 Rev. C / August 2016
Details
This command reads available data from the remote instrument and returns responses for the
specified number of arguments.
The format string can contain the following specifiers:
%[width]s
Read data until the specified length
%[max width]t
Read data until the specified length or until punctuation is found, whichever comes first
%[max width]n
Read data until a newline or carriage return
%d
Read a number (delimited by punctuation)
A maximum of 10 format specifiers can be used for a maximum of 10 return values.
If formatString is not provided, the command returns a string that contains the data until a new
line is reached. If no data is available, the Series 2600B pauses operation until the requested data is
available or until a timeout error is generated. Use tspnet.timeout to specify the timeout period.
When the Series 2600B reads from a TSP-enabled remote instrument, the Series 2600B removes
Test Script Processor (TSP
®
) prompts and places any errors it receives from the remote instrument
into its own error queue. The Series 2600B prefaces errors from the remote device with "Remote
Error," followed by the error number and error description.
Example
tspnet.write(deviceID, "*idn?\r\n")
print("write/read returns:", tspnet.read(deviceID))
Send the "*idn?\r\n" message to
the instrument connected as
deviceID.
Display the response that is read from
deviceID (based on the *idn?
message).
Also see
tspnet.connect() (on page 7-400)
tspnet.readavailable() (on page 7-404)
tspnet.timeout (on page 7-406)
tspnet.write() (on page 7-410)
tspnet.readavailable()
This function checks to see if data is available from the remote device.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
bytesAvailable = tspnet.readavailable(connectionID)
bytesAvailable
The number of bytes available to be read from the connection
connectionID
The connection ID returned from tspnet.connect()
Details
This command checks to see if any output data is available from the device. No data is read from the
instrument. This allows TSP scripts to continue to run without waiting on a remote command to finish.