EasyManua.ls Logo

Keithley 2600B

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-33
Example 2
printbuffer(1, 6, smua.nvbuffer1.sourcevalues)
Print the sourced value of the first 6 readings
stored in source-measure unit (SMU) A,
buffer 1.
Example output:
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04
Also see
bufferVar.measurefunctions (on page 7-25)
bufferVar.measureranges (on page 7-26)
bufferVar.n (on page 7-27)
bufferVar.readings (on page 7-28)
bufferVar.sourcefunctions (on page 7-29)
bufferVar.sourceoutputstates (on page 7-30)
bufferVar.sourceranges (on page 7-31)
bufferVar.statuses (on page 7-33)
bufferVar.timestamps (on page 7-35)
Reading buffers (on page 3-6)
bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Clearing the buffer
See Details
Not applicable
Usage
statusInformation = bufferVar.statuses[N]
statusInformation
The status value when reading N of the specified buffer was acquired
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
N
The reading number (
1
to bufferVar
.n
)

Table of Contents

Related product manuals