Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-412 2600BS-901-01 Rev. C / August 2016
Example 2
for name in userstring.catalog() do
print(name .. " = " ..
userstring.get(name))
end
Prints all userstring key-value pairs.
Output:
product = Widgets
assetnumber = 236
contact = John Doe
The above output lists the user-defined strings
added in the example for the userstring.add()
function. Notice the key-value pairs are not listed in
the order they were added.
Also see
userstring.add() (on page 7-410)
userstring.delete() (on page 7-412)
userstring.get() (on page 7-413)
userstring.delete()
This function deletes a user-defined string from nonvolatile memory.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
userstring.delete(name)
name
The name (key) of the key-value pair of the user-defined string to delete
Details
This function deletes the string that is associated with name from nonvolatile memory.
Example
userstring.delete("assetnumber")
userstring.delete("product")
userstring.delete("contact")
Deletes the user-defined strings associated with the
assetnumber, product, and contact names.
Also see
userstring.add() (on page 7-410)
userstring.catalog() (on page 7-411)
userstring.get() (on page 7-413)