Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-25
Example
smua.nvbuffer1.fillmode = smua.FILL_ONCE
Sets fill mode of dedicated reading
buffer 1 (source-
measure unit (SMU)
channel A) to fill once (do not
overwrite old data).
Also see
bufferVar.fillcount (on page 7-23)
Reading buffers (on page 3-6)
bufferVar.measurefunctions
This attribute contains the measurement function that was used to acquire a reading stored in a specified reading
buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Clearing the buffer
See Details
Not applicable
Usage
measurefunction = bufferVar.measurefunctions[N]
measurefunction
The measurement function used ("current", "voltage", "ohms", or "watts")
to acquire reading number
N
in the specified buffer
bufferVar
The reading buffer; can be a dynamically allocated buffer (user-defined), or a
dedicated reading buffer (such as smua.nvbuffer1)
N
The reading number (1 to bufferVar.n)