512 Keysight N9927-90001 User’s Guide
AM/FM Metrics (SA Only and Option 355)
Using AM/FM Metrics
How to Set Up AM/FM Metrics
Connect a test tone to the FieldFox SA Port 2 (Port 2)
—Press Measure
—Then Analog Demod / Tune & Listen
Then choose one of the following:
AM Metrics
or
FM Metrics
Left display - Frequency Spectrum (Power vs. Frequency)
Right display - Demodulated waveform vs. time where Demod SwpT
sets
the width of the demod window.
—Then Freq/Dist
> Center to set the carrier frequency.
—Set Freq Span
so that all sidebands above -60 dB are displayed in the left
window.
Optionally, adjust the Res BW
to better identify side bands.
Metrics are displayed in the AM/FM Modulation Metrics table on the
FieldFox. For AM and FM Metrics examples, refer to Figure 25-1 through
Figure 25-8 on page 521.
Selecting FM Wide vs. FM Narrow does not effect the displayed waveforms
or metrics. But, if you are using the Audio Demod feature and listening to the
demodulated waveform, then select the appropriate FM type, because the
tones do differ based on the FM type selected. Refer to Table 25-1 on
page 510.

Table of Contents

Other manuals for Keysight Technologies FieldFox N9938A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies FieldFox N9938A and is the answer not in the manual?

Keysight Technologies FieldFox N9938A Specifications

General IconGeneral
BrandKeysight Technologies
ModelFieldFox N9938A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals