EasyManuals Logo

Keysight Technologies FieldFox N9938A User Manual

Keysight Technologies FieldFox N9938A
756 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #519 background imageLoading...
Page #519 background image
Keysight N9927-90001 User’s Guide 519
AM/FM Metrics (SA Only and Option 355)
Interpreting AM/FM Metrics Error Messages
Interpreting AM/FM Metrics Error Messages
This section discusses possible AM/FM Metrics error messages and how to
correct them.
Test Tone Not Found
A Test Tone Not Found message is displayed under the following
circumstances:
FieldFox frequency span is too wide
No side lobes are displayed in the results
Refer to the examples in “Test Tone Not Found - Example” on page 520 and
“Test Tone Not Found - Corrected Example” on page 521.