SENSe:SWEep:OFFSet:POINts <value>|MIN|MAX
SENSe:SWEep:OFFSet:POINts? [MIN|MAX]
Defines the offset in a data sweep for triggered measurements. Positive values represent the delay
after the trigger occurs but before the samples are acquired. Negative values represent data samples
taken prior to the trigger.
Parameter Typical Return
–524,287 to 2,000,000,000, *RST 0 <offsetpoints>
Specifies -2048 offsetpoints: SENS:SWE:OFFS:POIN -2048
SENSe:SWEep:POINts <value>|MIN|MAX
SENSe:SWEep:POINts? [MIN|MAX]
Defines the number of points in a measurement. The number of points depends on the line frequency
(50 Hz or 60 Hz). The default number of points result in a measurement of 1 NPLC.
Parameter Typical Return
1 to 524,288 MIN|MAX,
*RST 3255 (60 Hz); 3906 (50 Hz)
<points>
Specifies 2048 points: SENS:SWE:POIN 2048
SENSe:SWEep:TINTerval <value>|MIN|MAX
SENSe:SWEep:TINTerval? [MIN|MAX]
Defines the time period between measurement samples. Units are in seconds. Values are rounded to
the nearest 20.48 microsecond increment. Below 20.48 microseconds>, values are rounded to the
nearest 10.24 or 5.12 microsecond increment respectively.
Parameter Typical Return
0.00000512 to 40,000, *RST 0.00000512 <time interval
Specifies an interval of 1 ms between points: SENS:SWE:TINT0.001
SENSe:THReshold<1|2|3|4>:FUNCtion AHOur|CURRent|POWer|VOLTage|WHOur
SENSe:THReshold<1|2|3|4>:FUNCtion?
Sets the sensing function for comparators THR1, THR2, THR3 or THR4.
AHOur selects the amp-hour function.
CURRent selects the current function.
POWer selects the power function.
VOLTage selects the voltage function.
WHOur selects the watt-hour function.
Keysight N6900/N7900 Series Operating and Service Guide 260
6 SCPI Programming Reference