OUTPut Subsystem
The Output subsystem controls the output state, power-on, protection, and relay functions.
OUTPut
[:STATe] 0|OFF|1|ON Enables or disables the output.
:COUPle
[:STATe] 0|OFF|1|ON Enables or disables output coupling.
:DOFFset <value> Sets a delay offset to synchronize coupled output state
changes.
:MAX
:DOFFset? Returns the delay offset required for this instrument.
:OFF
:SOURce EXPRession<1-8>|NONE Sets the off-couple signal source to an expression.
:ON
:SOURce EXPRession<1-8>|NONE Sets the on-couple signal source to an expression.
:DELay
:FALL <value> Sets the output turn-off sequence delay.
:RISE <value> Sets the output turn-on sequence delay.
:INHibit
:MODE LATChing|LIVE|OFF Sets the operating mode of the remote inhibit digital
pin.
:PON
:STATe RST|RCL0 Sets the output power-on state.
:PROTection
:CLEar Resets the latched protection.
:MODE Sets the turn-off behavior for all protection conditions.
:TEMPerature
:MARGin? Returns the margin remaining before the over-tem-
perature trips.
:USER
[:STATe] 0|OFF|1|ON Enables or disables the user-defined protection.
:SOURce EXPRession<1-8>|NONE Sets the user-defined protection source to an expres-
sion.
:WDOG
[:STATe] 0|OFF|1|ON Enables or disables the I/O watchdog timer.
:DELay <value> Sets the watchdog delay time.
:RELay
:LOCK
[:STATe] 0|OFF|1|ON Enables or disables the locked-closed state of the out-
put relays.
:POLarity NORMal|REVerse Sets the polarity of the output relays.
:ENABle 0|OFF|1|ON Enables or disables the polarity reversal function.
6 SCPI Programming Reference
247 Keysight N6900/N7900 Series Operating and Service Guide