Test Record Forms
N6950A/N6970A
N6950A/N6970A Test Record Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
+ 7.5 mV
Vout – 1.5 mV
8.9958 V
Vout – 4.2 mV
_______
_______
_______
_______
+ 10.5 mV
Vout + 1.5 mV
9.0042 V
Vout + 4.2 mV
CV Load Effect: Both – 0.5 mV _______ + 0.5 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 9 mV
+ 1 mV
Transient response @ 100 µs: Both – 150 mV _______ + 150 mV
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
N6950A
N6970A
N6950A
N6970A
N6950A
N6970A
N6950A
N6970A
– 30 mA
– 60 mA
Iout – 30 mA
Iout – 60 mA
99.87 A
199.74 A
Iout – 0.13 A
Iout – 0.26 A
_______
_______
_______
_______
_______
_______
_______
_______
+ 30 mA
+ 60 mA
Iout + 30 mA
Iout + 60 mA
100.13 A
200.26 A
Iout + 0.13 A
Iout + 0.26 A
CC Load Effect: N6950A
N6970A
– 8 mA
– 15 mA
_______
_______
+ 8 mA
+ 15 mA
Current Sink Tests
10% of current rating:
100% of current rating:
N6950A
N6970A
N6950A
N6970A
– 10.04 A
– 20.08 A
– 100.13 A
– 200.26 A
_______
_______
_______
_______
– 9.96 A
– 19.92 A
– 99.87 A
– 199.74 A
N6950A Settings N6970A Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
10% of current rating (Isink):
100% of current rating (Isink):
Voltage priority: 9 mV, 100 A
Voltage priority: 9 V, 100 A
Voltage priority: 9 V, 100 A
Voltage priority: 9 V, 50 A to 100 A
Current priority: 9 V,0 A
Current priority: 9 V,100 A
Current priority: 9 V,100 A
Current priority: 9 V,- 10 A
External supply: 9 V, 11 A
Current priority: 9 V,- 100 A
External supply: 9 V, 110 A
Voltage priority: 9 mV, 200 A
Voltage priority: 9 V, 200 A
Voltage priority: 9 V, 200 A
Voltage priority: 9 V, 100 A to 200 A
Current priority: 9 V,0 A
Current priority: 9 V,200 A
Current priority: 9 V,200 A
Current priority: 9 V,- 20 A
External supply: 9 V, 22 A
Current priority: 9 V,- 200 A
External supply: 9 V, 220 A
7 Verification and Calibration
322 Keysight N6900/N7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N6900 Series and is the answer not in the manual?

Keysight Technologies N6900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelN6900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals