2 kW Specifications N6970A N6971A N6972A N6973A N6974A
Voltage programming &
measurement accuracy
4, 5, 6
Lead drop ≤1 V max:
Lead drop ≤25% of V rating:
Lead drop ≤1 V max:
Lead drop ≤25% of V rating:
0.03% +1.5 mV
0.03% +1.9 mV
0.03% +1 mV
0.03% +1.4 mV
0.03% +3 mV
0.03% +4 mV
0.03% +2 mV
0.03% +3 mV
0.03% +6 mV
0.03% +7.9 mV
0.03% +4 mV
0.03% +5.9mV
0.03% +9 mV
0.03% +12 mV
0.03% +6 mV
0.03% +9 mV
0.03% +12 mV
0.03% +16 mV
0.03% +8 mV
0.03% +12mV
Current programming &
measurement accuracy:
4
accuracy:
0.1% +60 mA
0.04% +30 mA
0.1% +30 mA
0.04% +15 mA
0.1% +15 mA
0.04% +8 mA
0.1% +10 mA
0.04% +5 mA
0.1% +8 mA
0.04% +4 mA
Current low range
measurement accuracy:
0.05% +6 mA 0.05% +2 mA 0.05% +1.2 mA 0.05% +0.6 mA 0.05%+0.5 mA
Transient response
7
Recovery Time:
Settling band:
100 µs
150 mV
100 µs
150 mV
100 µs
100 mV
100 µs
150 mV
100 µs
200 mV
1
Current sinking up to 100% of rated current requires one power dissipator for 1 kW models, and two power dissipators for 2 kW models. 2
kW models with one power dissipator can sink 50% of their rated current.
2
Maximum continuous power available is derated at 1% of rating per degree C from 40°C to 55°C
3
From 20 Hz to 300 kHz for rms noise; from 20 Hz to 20 MHz for peak-to-peak noise
4
At 23°C ±5°C after a 30 minute warm-up; measurement NPLC=1; valid for 1 year, see Calibration Interval
5
With resistance programming enabled, the voltage programming offset component increases by a factor of 2.1
6
Load lead drop applies to each load lead
7
Time to recover to within the settling band following a load change from 50% to 100% of full load (10 µs rise time)
Specifications - Keysight N7900 Series
1 kW Specifications N7950A N7951A N7952A N7953A N7954A
DC Ratings
Voltage source:
Current source:
Current sink 10% rating:
Current sink 100% rating:
1
Power:
2
0 to 9 V
0 to 100 A
-10 A
-100 A
900 W
0 to 20 V
0 to 50 A
-5 A
-50 A
1 kW
0 to 40 V
0 to 25 A
-2.5 A
-25 A
1 kW
0 to 60 V
0 to 16.7 A
-1.67 A
-16.7 A
1 kW
0 to 80 V
0 to 12.5 A
-1.25 A
-12.5 A
1 kW
Output ripple & noise
3
CV rms:
CV peak-to-peak:
1 mV
9 mV
Load regulation
Voltage:
Current:
0.5 mV
8 mA
0.75 mV
3 mA
1.5 mV
1 mA
2 mV
1 mA
2 mV
0.8 mA
Voltage programming &
measurement accuracy
4, 5, 6
Lead drop ≤1 V max:
Lead drop ≤25% of V rating:
0.03% +1 mV
0.03% +1.4 mV
0.03% +2 mV
0.03% +3 mV
0.03% +4 mV
0.03% +5.9mV
0.03% +6 mV
0.03% +9 mV
0.03% +8 mV
0.03% +12mV
1 Quick Reference
42 Keysight N6900/N7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N6900 Series and is the answer not in the manual?

Keysight Technologies N6900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelN6900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals