STEP Command
[SOURce:]STEP:TOUTput 0|OFF|1|ON
[SOURce:]STEP:TOUTput?
Specifies whether a trigger out is generated when a transient step occurs. A trigger is generated when
the state is on (true).
Parameter Typical Return
0|OFF|1|ON , *RST OFF 0or1
Sets the step trigger signal to ON: STEP:TOUTON
6 SCPI Programming Reference
269 Keysight N6900/N7900 Series Operating and Service Guide