:DEFine EXPRession<1-8>,
<"expression">
Defines up to eight signal expressions.
:TEMPerature
:AMBient? Returns the temperature at the air inlet.
:TIME <hh>,<mm>,<ss> Sets the time of the system clock.
:VERSion? Returns the SCPI version that the instrument complies
with.
TRIGger Subsystem
TRIGger
:ACQuire
[:IMMediate] Generates an immediate trigger.
:CURRent
[:LEVel] <value> Sets the triggered level of the output.
:SLOPe POSitive|NEGative Sets the slope of the signal.
:INDices
[:DATA]? Returns the indices where triggers were captured.
:COUNt? Returns the number of triggers captured during the acquisition.
:SOURce <source> Selects the trigger source for the acquisition system: BUS |CURRent1
|EXTernal |EXPRession<1-8> |PIN<1-7> |TRANsient1 |VOLTage1
:TOUTput
[:ENABle] 0|OFF|1|ON Enables measurement triggers to be sent to a digital port pin.
:VOLTage
[:LEVel] <value> Sets the triggered level of the output.
:SLOPe POSitive|NEGative Sets the slope of the signal.
:ARB
:SOURce <source> Selects the trigger source for arbitrary waveforms: BUS |EXTernal
|IMMediate |EXPRession<1-8> |PIN<1-7>
:ELOG
[:IMMediate] Generates an immediate trigger.
:SOURce <source> Selects the trigger source for external data logging: BUS |EXTernal
|IMMediate| PIN<1-7>
:TRANsient
[:IMMediate] Generates an immediate trigger.
:SOURce <source> Selects the trigger source for the transient system: BUS |EXTernal
|IMMediate |EXPRession<1-8> |PIN<1-7>
1 Quick Reference
36 Keysight N6900/N7900 Series Operating and Service Guide