MEASure Subsystem
Measure commands measure the output voltage or current. They trigger the acquisition of new data
before returning the reading. Measurements are performed by digitizing the instantaneous output
voltage or current for a specified measurement time, storing the results in a buffer, and calculating the
value for the specified measurement type.
MEASure
[:SCALar]
:CURRent
[:DC]? Takes a measurement; returns the averaged current.
:ACDC? Takes a measurement; returns the RMS current (AC + DC).
:HIGH? Takes a measurement; returns the High level of a current pulse.
:LOW? Takes a measurement; returns the Low level of a current pulse.
:MAXimum? Takes a measurement; returns the maximum current.
:MINimum? Takes a measurement; returns the minimum current.
:POWer
[:DC]? Takes a measurement; returns the averaged power.
:MAXimum? Takes a measurement; returns the maximum power.
:MINimum? Takes a measurement; returns the minimum power.
:VOLTage
[:DC]? Takes a measurement; returns the averaged voltage.
:ACDC? Takes a measurement; returns the RMS voltage (AC + DC).
:HIGH? Takes a measurement; returns the High level of a voltage pulse.
:LOW? Takes a measurement; returns the Low level of a voltage pulse.
:MAXimum? Takes a measurement; returns the maximum voltage.
:MINimum? Takes a measurement; returns the minimum voltage.
:ARRAY
:CURRent
[:DC]? Takes a measurement; returns the instantaneous current.
:POWer
[:DC]? Takes a measurement; returns the instantaneous power.
:VOLTage
[:DC]? Takes a measurement; returns the instantaneous voltage.
MEASure[:SCALar]:CURRent[:DC]?
MEASure[:SCALar]:VOLTage[:DC]?
MEASure[:SCALar]:POWer[:DC]?
Initiates, triggers, and returns the averaged output measurement. Values returned are either in
amperes, volts, or watts.
Keysight N6900/N7900 Series Operating and Service Guide 244
6 SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N6900 Series and is the answer not in the manual?

Keysight Technologies N6900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelN6900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals