Page #324 background imageLoading...
Page #324 background image
N6952A/N6972A
N6952A/N6972A Test Record Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
+ 34 mV
Vout – 6 mV
39.982 V
Vout – 18 mV
_______
_______
_______
_______
+ 46 mV
Vout + 6 mV
40.018 V
Vout + 18 mV
CV Load Effect: Both – 1.5 mV _______ + 1.5 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 9 mV
+ 1 mV
Transient response @ 100 µs: Both – 100 mV _______ + 100 mV
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
N6952A
N6972A
N6952A
N6972A
N6952A
N6972A
N6952A
N6972A
– 8 mA
– 15 mA
Iout – 8 mA
Iout – 15 mA
24.967 A
49.935 A
Iout – 0.033 A
Iout – 0.065 A
_______
_______
_______
_______
_______
_______
_______
_______
+ 8 mA
+ 15 mA
Iout + 8 mA
Iout + 15 mA
25.033 A
50.065 A
Iout + 0.033 A
Iout + 0.065 A
CC Load Effect: N6952A
N6972A
– 1 mA
– 1.5 mA
_______
_______
+ 1 mA
+ 1.5 mA
Current Sink Tests
10% of current rating:
100% of current rating:
N6952A
N6972A
N6952A
N6972A
– 2.5105 A
– 5.02 A
– 25.033 A
– 50.065 A
_______
_______
_______
_______
– 2.4895 A
– 4.98 A
– 24.967 A
– 49.935 A
N6952A Settings N6972A Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
10% of current rating (Isink):
100% of current rating (Isink):
Voltage priority: 40 mV, 25 A
Voltage priority: 40 V, 25 A
Voltage priority: 40 V, 25 A
Voltage priority: 40 V, 12.5 A to 25 A
Current priority: 40 V, 0 A
Current priority: 40 V, 25 A
Current priority: 40 V, 25 A
Current priority: 40 V, - 2.5 A
External supply: 40 V, 2.75 A
Current priority: 40 V, - 25 A
External supply: 40 V, 27.5 A
Voltage priority: 40 mV, 50 A
Voltage priority: 40 V, 50 A
Voltage priority: 40 V, 50 A
Voltage priority: 40 V, 25 A to 50 A
Current priority: 40 V, 0 A
Current priority: 40 V, 50 A
Current priority: 40 V, 50 A
Current priority: 40 V, - 5 A
External supply: 40 V, 5.5 A
Current priority: 40 V, - 50 A
External supply: 40 V, 55 A
7 Verification and Calibration
324 Keysight N6900/N7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N6900 Series and is the answer not in the manual?

Keysight Technologies N6900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelN6900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals