:DATA <value> Sets the state of the digital control port.
:PIN<1-7>
:FUNCtion <function> Sets the function of the pins. DIO |DINPut |EXPRession<1-8>
|FAULt |INHibit |ONCouple |OFFCouple |TOUTput |TINPut
:POLarity POSitive|NEGative Sets the polarity of the pins.
:TOUTput
:BUS
[:ENABle] 0|OFF|1|ON Enables or disables BUS triggers on digital port pins.
[SOURce:]FUNCtion Command
[SOURce:]
:FUNCtion CURRent|VOLTage Sets the output regulation - voltage priority or current priority.
[SOURce:]LIST Subsystem
[SOURce:]
LIST
:COUNt <value>|INFinity Sets the list repeat count.
:CURRent
[:LEVel] <value>{,<value>} Specifies the setting for each list step.
:POINts? Returns the number of list points.
:DWELl <value>{,<value>} Specifies the dwell time for each list step.
:POINts? Returns the number of list points.
:STEP ONCE|AUTO Specifies how the list responds to triggers.
:TERMinate
:LAST 0|OFF|1|ON Determines the output value when the list terminates.
:TOUTput
:BOSTep
[:DATA] <Bool>{,<Bool>} Generates a trigger out at the Beginning Of STep
:POINts? Returns the number of list points.
:EOSTep
[:DATA] <Bool>{,<Bool>} Generates a trigger out at the End Of STep
:POINts? Returns the number of list points.
:VOLTage
[:LEVel] <value>{,<value>} Specifies the setting for each list step.
:POINts? Returns the number of list points.
[SOURce:]POWer Query
[SOURce:]
:POWer
:LIMit? Returns the output power limit of the instrument.
Keysight N6900/N7900 Series Operating and Service Guide 33
1 Quick Reference