SENSe:WINDow[:TYPE] HANNing|RECTangular
SENSe:WINDow[:TYPE]?
Selects the measurement window. This sets a signal conditioning function used in scalar DC
measurement calculations. Neither window function alters the instantaneous voltage or current data
returned in the measurement array.
Hanning window is a "raised cosine" function. It is a signal conditioning function that reduces errors in
DC measurement calculations in the presence of periodic signals such as AC line ripple. This window
only works up to 4883 measurement points. The instrument will revert to a rectangular window when
the points exceed 4883.
Rectangular window returns measurement calculations with no signal conditioning.
Parameter Typical Return
HANNing|RECTangular,
*RST RECTangular
RECT or HANN
Specifies a Hanning window function: SENS:WIND HANN
Keysight N6900/N7900 Series Operating and Service Guide 262