9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
"Questionable Temperature Event Query" on page 1125
"Questionable Temperature Negative Transition" on page 1125
"Questionable Temperature Positive Transition" on page 1125
Questionable Temperature Condition
Returns the decimal value of the sum of the bits in the Questionable Temperature
Condition register.
NOTE
The data in this register is continuously updated and reflects the current
conditions.
Remote Command
:STATus:QUEStionable:TEMPerature:CONDition?
Example
:STAT:QUES:TEMP:COND?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Temperature Enable
Determines which bits in the "Questionable Temperature Condition" on page 1124
Register will set bits in the "Questionable Temperature Event Query" on page 1125
register, which also sets the Temperature Summary bit (bit 4) in the "Questionable
Register" on page 1117.
The variable <integer> is the sum of the decimal values of the bits you want to
enable.
Remote
Command
:STATus:QUEStionable:TEMPerature:ENABle <integer>
:STATus:QUEStionable:TEMPerature:ENABle?
Example
:STAT:QUES:TEMP:ENAB 2
Instrument Overheated will be reported to the Temperature Summary of the Questionable Register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
IQ Analyzer Mode User's &Programmer's Reference 1124